Concepedia

Publication | Open Access

Small‐field beam data acquisition, detector dependency, and film‐based validation for a novel self‐shielded stereotactic radiosurgery system

12

Citations

29

References

2021

Year

Abstract

Minimal differences were seen in OAR and PDD measurements acquired with the diodes and the µD. ROFs measured with the three diodes were within 2.6% and within 1.4% versus the µD. Gafchromic Film measurements provided independent verification of the OAR and OF measurements. Estimated corrections to OFs using film as a reference were <1.6% for the Diode E, EDGE, and µD detector.

References

YearCitations