Influence of anterior capsulorhexis shape, centration, size, and location on intraocular lens position: finite element model
19
Citations
30
References
2021
Year
Anterior capsulorhexis irregularity and/or eccentricity increase IOL tilt, decentration, rotation, and vaulting in a numerically significant but optically negligible way. Von Mises stress is much greater at the capsular bag equator compared with the rhexis edge and highly asymmetrically distributed in all cases. Stress asymmetry may influence postoperative biologic processes of capsular bag shrinking and further IOL tilting or decentration.