Predictability of Refractive Outcome of a Small-Aperture Intraocular Lens in Eyes With Irregular Corneal Astigmatism

15

Citations

17

References

2021

Year

Abstract

The IC-8 IOL is well suited for patients with lens exchange in highly irregular corneas. The Haigis formula seemed to be the most accurate in the patient group. Ray-tracing confirmed the results of biometric formulas up to a keratometric value of 48.00 D and should be compared with standard biometric formulas to address corneal irregularities and to minimize refractive surprises after surgery. A comparison with ray-tracing in eyes with a keratometric value of greater than 48.00 D should not be considered due to the inaccurate results. <b>[<i>J Refract Surg</i>. 2021;37(5):312-317.]</b>.