Anaesthesia and Intensive Care · 2021 · 20 citations · 18 references
Health ProfessionsClinical SciencesRespiratory TherapyCritical Care Respiratory TherapyPulmonologyVentilator ManagementAcute Lung InjuryMechanical VentilationPatient-ventilator SynchronyPositive End-expiratory PressureRespiratory Failure ManagementVentilatory SupportCycle AsynchronyDelayed CyclingMechanical Ventilation ManagementPremature CyclingVentilation OptimisationVentilator AsynchronyVentilator MonitoringVentilator-induced Asynchrony
Patient-ventilator dyssynchrony or asynchrony occurs when, for any parameter of respiration, discordance exists between the patient's spontaneous effort and the ventilator's provided support. If not recognised, it may promote oversedation, prolong the duration of mechanical ventilation, create risk for lung injury, and generally confuse the clinical picture. Seven forms of dyssynchrony are common: (a) ineffective triggering; (b) autotriggering; (c) inadequate flow; (d) too much flow; (e) premature cycling; (f) delayed cycling; and (g) peak pressure apnoea. 'Reverse triggering' also occurs and may mimic premature cycling. Correct diagnosis of these phenomena often permits management by simple ventilator optimisation rather than by less desirable measures.
18
Patient-ventilator asynchrony during assisted mechanical ventilation
Arnaud W. Thille, Pablo Rodríguez, Belén Cabello et al. · Intensive Care Medicine · 2006 · 923 citations · Full text
Mechanical Ventilation-Induced Reverse-Triggered Breaths
Evangelia Akoumianaki, Aissam Lyazidi, Nathalie Rey et al. · CHEST Journal · 2013 · 294 citations