Peak Expiratory Flow During Mechanical Insufflation-Exsufflation: Endotracheal Tube Versus Face Mask

Sung Eun Hyun, Sang‐Min Lee, Hyung‐Ik Shin

Respiratory Care · 2021 · 14 citations · 23 references

DOIFull text

Open access

Abstract

MI-E via ETT generated slower PEF than via face mask, suggesting that a higher-pressure protocol should be prescribed for intubated patients. An insufflation-exsufflation pressure up to +50/-50 cm H<sub>2</sub>O could be considered to produce a PEF faster than 2.7 L/s, and the applications were safe and feasible for subjects on invasive mechanical ventilation.

References

23