PubMed · 2021 · 14 citations · 24 references
DentistryImplant DentistryDigital Implant DentistryDigital ProsthodonticsImplant ImagingDental ImagingDigital Dental ImagingImplant Impression ScanningImplant-abutment InterfaceIn-vitro CharacterizationIntraoral Scan AccuracyPrecisionScanner ValidationScanning AccuracyScanning Precision
STATEMENT OF THE PROBLEM: The mechanisms of operation of dental scanners are based on different technologies. Considering these differences, there are many types of scanners available in the market. PURPOSE: This in vitro study aimed to compare the accuracy (precision and trueness) of seven commonly used dental scanners. MATERIALS AND METHOD: In this in vitro experimental study, the accuracy of 7 common extra oral scanners (Sirona ineos inLab, Sirona X5, Dentium, Imes Icore 350I, Amann Girrbach, 3shape D700, and 3Shape E3) were evaluated. Each of scanners performed 7 scans of implant abutment of SIC (SIC MAX.GH1). Data from each scanner were then compared to data received from 3Shape Trios intra oral scanner as a reference. For evaluating the accuracy of each scanner, trueness and precision was evaluated. Collected data were analyzed using Kruskal Wallis and Bonferroni tests via SPSS version 22. RESULTS: <0.05). CONCLUSION: Based on the results of this study, the extra oral scanner, 3shape E3, had the best trueness and precision. The lowest amount of trueness among the studied scanners was for the extra oral scanner, Sirona x5, and the lowest precision was for scanner 3shape D700.
24