Scientific Reports · 2021 · 110 citations · 24 references
EngineeringBioengineeringBiomedical EngineeringBiomedical Signal ProcessingElectromagnetismElectrophysiologyCardiac Signal ProcessingCardiovascular MonitoringElectrocardiogram AnalysisElectrocardiographyMagnetic SensorsCardiac Monitoring SystemsEcg Signal DenoisingElectrocardiogram Feature ExtractionElectromagnetic ImagingEmpirical Mode DecompositionIndependent Component AnalysisMultichannel Data AcquisitionSpin Exchange
In this study, a wearable multichannel human magnetocardiogram (MCG) system based on a spin exchange relaxation-free regime (SERF) magnetometer array is developed. The MCG system consists of a magnetically shielded device, a wearable SERF magnetometer array, and a computer for data acquisition and processing. Multichannel MCG signals from a healthy human are successfully recorded simultaneously. Independent component analysis (ICA) and empirical mode decomposition (EMD) are used to denoise MCG data. MCG imaging is realized to visualize the magnetic and current distribution around the heart. The validity of the MCG signals detected by the system is verified by electrocardiogram (ECG) signals obtained at the same position, and similar features and intervals of cardiac signal waveform appear on both MCG and ECG. Experiments show that our wearable MCG system is reliable for detecting MCG signals and can provide cardiac electromagnetic activity imaging.
24
A subfemtotesla multichannel atomic magnetometer
I. K. Kominis, T. W. Kornack, Joel C. Allred et al. · Nature · 2003 · 1.6K citations
High-Sensitivity Atomic Magnetometer Unaffected by Spin-Exchange Relaxation
J. C. Allred, ROBERT LYMAN, T. W. Kornack et al. · Physical Review Letters · 2002 · 1K citations