Japanese Journal of Applied Physics · 2021 · 20 citations · 34 references
EngineeringBioengineeringBiomedical EngineeringBiomemsBiosensorsMicroelectromechanical SystemsBiomedical MemsBiosensor DesignBiosensors For Immune DetectionMems ResonatorsMicroelectromechanical SensorsMicroscale BiosensorsDevice MiniaturizationImmunosensorsMiniaturizationPower ConsumptionQuartz Crystal MicrobalanceQuartz Crystal Microbalance BiosensorsQuartz Crystal Microbalance ImmunosensorsQuartz Crystal Microbalance Sensors
Abstract We present a wireless quartz crystal microbalance (QCM) biosensor fabricated using MEMS technology. The MEMS QCM biosensor contains a 125 MHz AT-cut quartz resonator embedded in the microchannel. Because of the compact design, the MEMS QCM biosensor is suitable for mass production and device miniaturization. We performed immunoglobulin G (IgG) detection measurements with different concentrations of IgG. The detection limit was 1 ng ml –1 or less, which is superior to that of the gold-standard surface plasma resonance method. Furthermore, we studied the binding affinity between protein A and IgG by studying the frequency response of the QCM biosensor. We found good agreement with reported values. Therefore, the presented MEMS QCM biosensor has the advantages of compactness, low cost, low power consumption, high sensitivity, and reliability.
34