Blood Pressure Monitoring · 2021 · 82 citations · 14 references
MedicineCardiologyBioinstrumentationPhysiatryPhysical ExaminationBlood Pressure ControlBlood Pressure Signal AnalysisCardiovascular AssessmentCardiovascular MonitoringMedical InstrumentationOptical InstrumentsAuscultatory Blood Pressure MeasurementBlood PressureBlood Pressure MeasurementBlood Pressure MonitoringBlood Pressure MonitorsCuffless Blood Pressure MeasurementNoninvasive Blood Pressure MeasurementValidation Study Design
OBJECTIVE: The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month. METHODS: In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject. RESULTS: Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP. CONCLUSION: After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 in the sitting position. The Aktiia Bracelet can be recommended for BP measurement in the adult population.
14
2018 ESC/ESH Guidelines for the management of arterial hypertension
Bryan Williams, Giuseppe Mancia, Wilko Spiering et al. · Journal of Hypertension · 2018 · 2.8K citations