Transition From Manual to Automated Planning and Delivery of Volumetric Modulated Arc Therapy Stereotactic Radiosurgery: Clinical, Dosimetric, and Quality Assurance Results

34

Citations

28

References

2021

Year

Abstract

HyperArc produces high-quality radiosurgical plans that are at least as good as mVMAT plans created by an expert manual planner with easier planning and more efficient delivery workflow. A less experienced planner can produce very high-quality radiosurgical plans even for patients with more than 10 targets. The use of a single-isocenter technique for multiple targets with no PTV margin did not compromise clinical outcomes, and 1-year local control for treated targets remained congruent with historical series.