Reliability and Validity of the Lactate Pro 2 Analyzer
42
Citations
13
References
2021
Year
Physical ActivityNeuromuscular CoordinationCardiometabolic RiskLp2 MeasurementsKinesiologyExerciseBioanalysisPhysical ExerciseApplied PhysiologyClinical ExerciseClinical ChemistrySport PhysiologyLaboratory MedicineHealth SciencesReliabilityPhysical FitnessClinical Exercise PhysiologyBioinstrumentationExercise ScienceLactate Pro 2PhysiologyExercise PhysiologyYsi 1500Athletic TrainingMedicineLactate Threshold
This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n = 258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L−1 for blood lactate (BLa) ≤4.0 mmol.L−1 to −1.6 to +1.4 mmol.L−1 for BLa >8.0 mmol.L−1, with minimal systematic bias. Despite LP2 measurements being ~1.42 times higher than YSI measurements, no significant differences (P > .05) were detected when quantifying load, heart rate, or oxygen consumption data at lactate threshold. The results support using different LP2 devices interchangeably, especially when BLa ≤ 8 mmol.L−1, and using the LP2 to identify training intensities at lactate threshold.