Sensors · 2021 · 77 citations · 30 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingDiagnostic ImagingUltrasonographyLight-tissue InteractionMedical ImagingOptical ImagingPhotoacoustic ImagingTissue Optical PropertiesTissue OpticsGround Truth VerificationGround-truth ValidationOptical PropertiesOxygen SaturationPhotoacoustic Imaging SystemsSmall Animal ImagingTissue Oxygen Saturation MappingTissue Oxygenation Measurement
Oxygen saturation imaging has potential in several preclinical and clinical applications. Dual-wavelength LED array-based photoacoustic oxygen saturation imaging can be an affordable solution in this case. For the translation of this technology, there is a need to improve its accuracy and validate it against ground truth methods. We propose a fluence compensated oxygen saturation imaging method, utilizing structural information from the ultrasound image, and prior knowledge of the optical properties of the tissue with a Monte-Carlo based light propagation model for the dual-wavelength LED array configuration. We then validate the proposed method with oximeter measurements in tissue-mimicking phantoms. Further, we demonstrate in vivo imaging on small animal and a human subject. We conclude that the proposed oxygen saturation imaging can be used to image tissue at a depth of 6-8 mm in both preclinical and clinical applications.
30
Photoacoustic imaging in biomedicine
Minghua Xu, Lihong V. Wang · Review of Scientific Instruments · 2006 · 2.7K citations · Full text
Multispectral Optoacoustic Tomography (MSOT) of Human Breast Cancer
Gaël Diot, Stephan Metz, Aurelia Noske et al. · Clinical Cancer Research · 2017 · 273 citations