Journal of Synchrotron Radiation · 2020 · 33 citations · 45 references
Veterinary MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsRadiographic ImagingX-ray OpticsComputed Tomography InstrumentationRotation Speed ControlTomographic AcquisitionTomographic RadiographyX-ray Computed TomographyX-ray Imaging PhysicsX-ray Imaging Systems
The temporal resolution of X-ray tomography, using a synchrotron radiation X-ray source, has been improved to millisecond order in recent years. However, the sample must be rotated at a speed of more than a few thousand revolutions per minute, which makes it difficult to control the environment around the sample. In this study, a high-speed rotation device has been developed, comprising two synchronized coaxial motors movable along the direction of the axis, which can stretch or compress the rotating sample. Using this device, tomograms of breaking rubber were successfully obtained at a temporal resolution of 10 ms.
45
David L. Donoho · IEEE Transactions on Information Theory · 2006 · 23.1K citations
Phase-contrast imaging using polychromatic hard X-rays
S. Wilkins, Timur E. Gureyev, D. Gao et al. · Nature · 1996 · 1.8K citations