Physics and Imaging in Radiation Oncology · 2020 · 17 citations · 19 references
MedicineNuclear MedicineRadiologyNuclear InstrumentationRadiation DosimetryTreatment PlanningAdaptive RadiotherapyImage-guided RadiotherapyMagnetic Resonance ImagingMedical Imaging InstrumentationPhantom DosimetryRadiation TherapyAcceleratorsAdaptive Radiotherapy DosimetryLinear AcceleratorsMagnetic ResonanceMagnetic Resonance Imaging Phantom DosimetryMagnetic Resonance PhysicsMagnetic Resonance-guided RadiotherapyReal-time Imaging
The introduction of real-time imaging by magnetic resonance guided linear accelerators (MR-Linacs) enabled adaptive treatments and gating on the tumor position. Different end-to-end tests monitored the accuracy of our MR-Linac during the first year of clinical operation. We report on the stability of these tests covering a static, adaptive and gating workflow. Film measurements showed gamma passing rates of 96.4% ± 3.4% for the static tests (five measurements) and for the two adaptive tests 98.9% and 99.99%, respectively (criterion 2%/2mm). The gated point dose measurements in the breathing phantom were 2.7% lower than in the static phantom.