Precision of handheld multispectral optoacoustic tomography for muscle imaging

Alexandra L. Wagner, Vera Danko, Anna Federle, Daniel Klett, David Simón, Jörg Jüngert, Michael Uder, Georg Schett, Markus F. Neurath, Joachim Woelfle,

Photoacoustics · 2020 · 43 citations · 31 references

DOIFull text

Open access

Concepts

Abstract

Photo-or optoacoustic imaging (OAI) allows quantitative imaging of target tissues. Using multi-wavelength illumination with subsequent ultrasound detection, it may visualize a variety of different chromophores at centimeter depth. Despite its non-invasive, label-free advantages, the precision of repeated measurements for clinical applications is still elusive. We present a multilayer analysis of n = 1920 imaging datasets obtained from a prospective clinical trial (NCT03979157) in n = 10 healthy adult volunteers. All datasets were analyzed for 13 single wavelengths (SWL) between 660 nm-1210 nm and five MSOT-parameters (deoxygenated/oxygenated/total hemoglobin, collagen and lipid) by a semi-automated batch mode software. Intraclass correlation coefficients (ICC) were good to excellent for intrarater (SWL: 0.82-0.92; MSOT-parameter: 0.72-0.92) and interrater reproducibility (SWL: 0.79-0.87; MSOT-parameter: 0.78-0.86), with the exception for MSOT-parameter lipid (interrater ICC: 0.56). Results were stable over time, but exercise-related effects as well as inter-and intramuscular variability were observed. The findings of this study provide a framework for further clinical OAI implementation.