Clinical Management and Pump Parameter Adjustment of the Control-IQ Closed-Loop Control System: Results from a 6-Month, Multicenter, Randomized Clinical Trial
21
Citations
10
References
2020
Year
<b><i>Background:</i></b> Data are limited on the need for and benefits of pump setting optimization with automated insulin delivery. We examined clinical management of a closed-loop control (CLC) system and its relationship to glycemic outcomes. <b><i>Materials and Methods:</i></b> We analyzed personal parameter adjustments in 168 participants in a 6-month multicenter trial of CLC with Control-IQ versus sensor-augmented pump (SAP) therapy. Preset parameters (BR = basal rates, CF = correction factors, CR = carbohydrate ratios) were optimized at randomization, 2 and 13 weeks, for safety issues, participant concerns, or initiation by participants' usual diabetes care team. Time in range (TIR 70-180 mg/dL) was compared in the week before and after parameter changes. <b><i>Results:</i></b> In 607 encounters for parameter changes, there were fewer adjustments for CLC than SAP (3.4 vs. 4.1/participant). Adjustments involved BR (CLC 69%, SAP 80%), CR (CLC 68%, SAP 50%), CF (CLC 44%, SAP 41%), and overnight parameters (CLC 62%, SAP 75%). TIR before and after adjustments was 71.2% and 71.3% for CLC and 61.0% and 62.9% for SAP. The highest baseline HbA<sub>1c</sub> CLC subgroup had the largest TIR improvement (51.2% vs. 57.7%). When a CR was made more aggressive in the CLC group, postprandial time >180 mg/dL was 43.1% before the change and 36.0% after the change. The median postprandial time <70 mg/dL before making CR less aggressive was 1.8%, and after the change was 0.7%. <b><i>Conclusions:</i></b> No difference in TIR was detected with parameter changes overall, but they may have an effect in higher HbA<sub>1c</sub> subgroups or following user-directed boluses, suggesting that changes may matter more in suboptimal control or during discrete periods of the day. Clinical Trials Registration number: NCT03563313.
| Year | Citations | |
|---|---|---|
Page 1
Page 1