Journal of Cardiovascular Electrophysiology · 2020 · 12 citations · 7 references
MedicineSurgeryAnesthesiaCardiothoracic SurgeryMicrosurgeryAblation TherapyElectrosurgical InstrumentsEndoscopic SurgeryEsophageal SurgeryMinimally Invasive SurgeryRetracting InstrumentsAblation SafetyAblation TechnologyEsophageal AnatomyEsophageal Injury PreventionEsophageal Mucosal InjurySurgical Ablation
Deviation of the esophagus during atrial fibrillation (AF) ablation can reduce esophageal injury. This study reports upon a novel esophageal retractor that utilizes vacuum suction and mechanical deflection to deviate the esophagus. The device was used in seven patients undergoing cryoballoon AF ablation. The esophagus was deviated 31.9 ± 4.4 mm to the right and 28.2 ± 5.9 mm to the left. Endoscopy at 4.4 ± 1.5 days postablation showed no esophageal injury. This study demonstrates the safe and effective deviation of the esophagus without a trailing edge with an esophageal retractor utilizing vacuum suction and mechanical deflection.
7
Valay Parikh, Vijay Swarup, Jacob D. Hantla et al. · Heart Rhythm · 2018 · 60 citations
Medicine, Cardiology, Surgery +11
Esophageal Deviation During Atrial Fibrillation Ablation
Rahul Bhardwaj, Aditi Naniwadekar, William Whang et al. · JACC. Clinical electrophysiology · 2018 · 59 citations
Patrick Müller, Karin Nentwich, Joachim Krug et al. · EP Europace · 2016 · 55 citations · Full text