EP Europace · 2020 · 33 citations · 31 references
MedicineCardiologySurgeryElectrophysiologyInterventional CardiologyInterventional RadiologyAblation TherapyCatheter AblationAblation PhysicsAblation TechnologyCatheter Electrode TechnologyElectrode - Skin InterfaceElectrode-skin InterfaceIntracardiac EchocardiographyNon-thermal AblationSurgical Ablation
AIMS: Irreversible electroporation (IRE) ablation is a non-thermal ablation method based on the application of direct current between a multi-electrode catheter and skin electrode. The delivery of current through blood leads to electrolysis. Some studies suggest that gaseous (micro)emboli might be associated with myocardial damage and/or (a)symptomatic cerebral ischaemic events. The aim of this study was to compare the amount of gas generated during IRE ablation and during radiofrequency (RF) ablation. METHODS AND RESULTS: In six 60-75 kg pigs, an extracorporeal femoral shunt was outfitted with a bubble-counter to detect the size and total volume of gas bubbles. Anodal and cathodal 200 J IRE applications were delivered in the left atrium (LA) using a 14-electrode circular catheter. The 30 and 60 s 40 W RF point-by-point ablations were performed. Using transoesophageal echocardiography (TOE), gas formation was visualized. Average gas volumes were 0.6 ± 0.6 and 56.9 ± 19.1 μL (P < 0.01) for each anodal and cathodal IRE application, respectively. Also, qualitative TOE imaging showed significantly less LA bubble contrast with anodal than with cathodal applications. Radiofrequency ablations produced 1.7 ± 2.9 and 6.7 ± 7.4 μL of gas, for 30 and 60 s ablation time, respectively. CONCLUSION: Anodal IRE applications result in significantly less gas formation than both cathodal IRE applications and RF applications. This finding is supported by TOE observations.
31
Fiorenzo Gaïta, Jean Leclercq, Burghard Schumacher et al. · Journal of Cardiovascular Electrophysiology · 2011 · 285 citations · Full text
Medicine, Cardiology, Surgery +15
Low vulnerability of the right phrenic nerve to electroporation ablation
Vincent J.H.M. van Driel, Kars Neven, Harry van Wessel et al. · Heart Rhythm · 2015 · 180 citations