How large is the periablational zone after radiofrequency and microwave ablation? Computer-based comparative study of two currently used clinical devices

Macarena Trujillo, Punit Prakash, Pegah Faridi, Aleksandar Radosevic, Sergio Curto, Fernando Burdı́o, Enrique Berjano

International Journal of Hyperthermia · 2020 · 15 citations · 37 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: To compare the size of the coagulation (CZ) and periablational (PZ) zones created with two commercially available devices in clinical use for radiofrequency (RFA) and microwave ablation (MWA), respectively. METHODS: < 2.1 were considered as the PZ (tissue that has undergone moderate sub-ablative hyperthermia). The ratio of PZ volume to CZ volume (PZ/CZ) was regarded as a measure of performance, since a low value implies achieving a large CZ while keeping the PZ small. RESULTS: for 60-100 W MWA, respectively). PZ/CZ for RFA were relatively high (65-69%), regardless of ablation time, while those for MWA were highly dependent on the duration (increase of up to 25% between 5 and 10 min) and on the applied power (smaller values as power was raised, 102% for 60 W vs. 81% for 100 W, both for 10 min). The lowest PZ/CZ across all settings was 56%, obtained with 100 W-5 min MWA. CONCLUSIONS: Although RFA creates smaller periablational zones than MWA, 100 W-5 min MWA provides the lowest PZ/CZ.