Bioengineering · 2020 · 12 citations · 12 references
MedicineBioengineeringBiomedical EngineeringBiomedical ImagingCardiothoracic SurgeryFinite Element AnalysisAortic ImagingAortic StenosisAortic SurgeryComputed TomographyImplant BiomechanicsTranscatheter Aortic Valve ReplacementAortic Root AnatomyAortic Root AssessmentAortic Root DiseaseAortic Tissue BiomechanicsAortic Valve ImagingCalcificationProcedure ComplicationsValve Calcification Assessment
Gated computed tomography (CT) might not adequately predict occurrence of post-implantation transcatheter aortic valve replacement (TAVR) complications in hostile aortic root as it would require a more complex integration of morphological, functional and hemodynamical parameters. We used a computational framework based on finite element analysis (FEA) to simulate patient-specific implantation. Application of biomechanical modelling using FEA to gated-CT was able to demonstrate the relation of the device with voluminous calcification, its consequent misalignment and a significant stent deformation. Use of FEA and other advanced computed predictive modelling techniques as an adjunct to CT scan could improve our understanding of TAVR, potentially predict complications and fate of the devices after implantation and inform patient-specific treatment.
12
Omar Khalique, Rebecca T. Hahn, Hemal Gada et al. · JACC: Cardiovascular Interventions · 2014 · 223 citations
Medicine, Cardiology, Surgery +13