Concepedia

Publication | Open Access

A spot test for determination of residual TBA levels in<sup>18</sup>F-radiotracers for human use using Dragendorff reagent

13

Citations

20

References

2020

Year

Abstract

When utilizing [<sup>18</sup>F]tetrabutylammonium fluoride ([<sup>18</sup>F]TBAF) in the synthesis of <sup>18</sup>F-labeled radiotracers for clinical positron emission tomography (PET) imaging, it is necessary to confirm that residual TBA levels in formulated doses do not exceed established specifications (≤2.6 mg per patient dose). Historically this has been accomplished using HPLC, but this is time consuming for short-lived PET radiotracers and limited by the need for expensive equipment. This motivated us to introduce a TLC spot test for determining residual TBA, and we have developed a new method which employs the Dragendorff reagent. Herein we report details of the TLC method and use it to quantify residual TBA in different formulations of 6-[<sup>18</sup>F]fluoro-DOPA.

References

YearCitations