Tertiary referral center. A total of 95 patients with cochlear implants from one manufacturer were included in this study. The lateral wall, the modiolar wall, and the cochlear implant electrode were segmented using OsiriX MD. For repositioning and reconstructing the respective contours and measuring distances, files were analyzed in MATLAB. The distances from the edge of each electrode contact to the cochlear walls showed no significant differences. But between the different contacts within each patient, there were significant differences. Around 180 degree insertion, electrodes start to get in contact with the lateral wall. The tip of the electrode array was always facing toward the modiolar wall independent of the length of the electrode. We established a method to analyze the position of electrodes within the cochlea.