IEEE Journal of Solid-State Circuits · 2020 · 55 citations · 29 references
MedicineBiomedical EngineeringAnalog ElectronicsUltrasonographyUltrasoundInstrumentation AmplifiersLow-noise Circuit DesignTransducer TechnologyUltrasonicsUltrasound ImagingUltrasound InstrumentationApplication-specific Integrated Circuit Design MethodologyBandwidthGainMiniaturized Ultrasound ProbesPower-efficient DesignTime Gain CompensationTransimpedance Amplifier DesignTransimpedance Amplifiers
This article presents a low-noise transimpedance amplifier (TIA) designed for miniature ultrasound probes. It provides continuously variable gain to compensate for the time-dependent attenuation of the received echo signal. This time-gain compensation (TGC) compresses the echo-signal dynamic range (DR) while avoiding imaging artifacts associated with discrete gain steps. Embedding the TGC function in the TIA reduces the output DR, saving power compared to prior solutions that apply TGC after the low-noise amplifier. The TIA employs a capacitive ladder feedback network and a current-steering circuit to obtain a linear-in-dB gain range of 37 dB. A variable-gain loop amplifier based on current-reuse stages maintains constant bandwidth in a power-efficient manner. The TIA has been integrated in a 64-channel ultrasound transceiver application-specific integrated circuit (ASIC) in a 180-nm BCDMOS process and occupies a die area of 0.12 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . It achieves a gain error below ±1 dB and a 1.7 pA/√Hz noise floor and consumes 5.2 mW from a ±0.9 V supply. B-mode images of a tissue-mimicking phantom are presented that show the benefits of the TGC scheme.
29
Single-chip CMUT-on-CMOS front-end system for real-time volumetric IVUS and ICE imaging
Gokce Gurun, Coşkun Tekeş, Jaime Zahorian et al. · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · 2014 · 173 citations · Full text
Veterinary Medicine, Bioengineering, Biomedical Engineering +17