All ADM kits tested were dose responsive within the therapeutic range and correlated well. The significance of observed low-range false positives and cross reactivity with infliximab in LISA-Tracker and Shikari kits is dependent on the indications received for testing in the laboratory. Anti-ADM ELISA kits produced varied results for spiked sera; however, they showed good precision. Inter-kit variability suggested that anti-ADM levels should be compared only when using the same method.