PubMed · 2020 · 21 citations · 26 references
Open access
NursingCardiologyPhysiologyBioinstrumentationClinical CardiologyMobile Health InformaticsCardiac AssessmentCardiovascular AssessmentCardiovascular MonitoringElectrocardiographyHeart Rate VariabilityMobile Health Data CollectionBland-altman PlotConsistency Of Measurement ResultsHeart Rate MonitoringHeart Rate Variability MeasurementMeasurement ConsistencyR-r Interval VariabilityRepeated Measures AnalysisValidation Study Design
BACKGROUND: ) were calculated to determine the degree of association between the two electronic devices. Two-way repeated measures analysis of variance and Bland-Altman analysis were used to calculate the measurement consistency and agreement, respectively, between the two methods. Effect size was also used to estimate the magnitude of the differences. RESULTS: = 0.999; Confidence Interval [CI] at 95%: 0.999-0.999; P < 0.0001) and during orthostatic challenge (r = 0.988; 95% CI: 0.988-0.989; P < 0.0001). A negative bias of -0.526 milliseconds (95% limits of agreement [LoA] from -4.319 to 3.266 milliseconds) was observed in supine position between ECG and the smartphone application. On the other hand, a positive bias of 0.077 milliseconds (95% LoA from -10.090 to 10.240 milliseconds) during the orthostatic challenge was observed. CONCLUSIONS: Our results cross-validated a freeware smartphone application with the ECG-derived reference condition for asymptomatic adults at rest in the supine position and during orthostatic challenge.
26
Task Force of the European Society of Cardiology the North A Electrophysiology · Circulation · 1996 · 16.9K citations