Radiation Protection Dosimetry · 2020 · 13 citations · 17 references
MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DosimetryAbsorbed DoseComputed TomographyDose CalculationPatient-specific DosimetryRadiation Dose MeasurementComputed Tomography Dose IndicesComputed Tomography Dose MeasurementDose-length ProductMedical Imaging Radiation DosePatient Dose MeasurementPatient Radiation Dose OptimizationRadiation Dose AssessmentRadiation Dose Estimation
A pilot study has concerned the most frequent computed tomography examinations (CT). This represents the first results based on actual survey for diagnostic reference levels (DRLs) establishment in Algeria. A total number of 2540 patients underwent this survey that has included the recording of CT parameters, computed tomography dose index (CTDIvol) and dose-length product of the head, thorax, abdomen, abdomen-pelvis (AP), lumbar spine (LS) and thorax-abdomen-pelvis (TAP) performed on standard patients. The proposed DRLs are 71 mGy/1282 mGy.cm for head, 16 mGy/555 mGy.cm for thorax, 18 mGy/671 mGy.cm for abdomen, 21 mGy/950 mGy.cm for AP, 36 mGy/957 mGy.cm for LS and 18 mGy/994 mGy.cm for TAP. The rounded 75th percentile seems to be higher in some examinations compared to the literature. Our findings confirm the need to optimise our practice. These results provide a starting point for institutional evaluation of CT radiation doses.
17
ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging
E. Vañó, D.L. Miller, C J Martin et al. · Annals of the ICRP · 2017 · 981 citations