EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationMicroelectronicsAnalog InstrumentationAnalog Integrated CircuitsAnalog IntegrationCalibrationElectrical MeasurementIntegrated Circuit DesignAnalog TestabilityCalibration Techniques For Analog CircuitsCharacteristic ImpedanceImpedance MeasurementImpedance MetrologyIn Situ Impedance MeasurementsPrecision MeasurementPrototype Evaluation
In this work, the accuracy of the MAX30001 integrated-circuit for impedance measurements is characterized using measurements of resistors and 2R-1C configurations with |Z| <; 200Ω in the frequency band from 1 kHz to 128 kHz. Measurements were collected using a prototype system that integrated analog multiplexors to enable measurements of up to 4 test impedances by routing the MAX30001 excitation/measurement signals to multiple sites without manual switching. The additional analog switching circuitry did not degrade the MAX30001 measurements. Using a two-point calibration for the MAX30001 reduced the error of resistance measurements from a maximum of 4.29% to <; 0.22%, but measurements of complex impedances showed significantly higher errors which require further investigation to identify the source of errors and implement methods to correct.