Evaluation of scatter rejection and correction performance of 2D antiscatter grids in cone beam computed tomography

18

Citations

34

References

2021

Year

Abstract

When 2D ASG is used solely as a scatter rejection device, substantial improvement in CT number accuracy can be achieved by increasing the grid ratio. Two-dimensional ASGs also provided significant CNR improvement even at lower grid ratios. Two-dimensional ASGs used in conjunction with the grid-based scatter sampling method provided further improvement in CT number accuracy, irrespective of the grid ratio, while preserving 2D ASGs' capacity to improve CNR. The combined effect of scatter rejection and residual scatter correction by 2D ASG may accelerate implementation of new techniques in CBCT that require high quantitative accuracy, such as radiotherapy dose calculation and dual energy CBCT.