Low-Noise Magnetic Coil System for Recording 3-D Eye Movements

Kristin N. Hageman, Margaret R. Chow, Dale Roberts, Charles C. Della Santina

IEEE Transactions on Instrumentation and Measurement · 2020 · 386 citations · 24 references

DOIFull text

Open access

Concepts

Abstract

OBJECTIVE: Vestibular and oculomotor research often requires measurement of 3-dimensional (3D) eye orientation and movement with high spatial and temporal precision and accuracy. We describe the design, implementation, validation and use of a new magnetic coil system optimized for recording 3D eye movements using small scleral coils in animals. METHODS: Like older systems, the system design uses off-the-shelf components to drive three mutually orthogonal alternating magnetic fields at different frequencies. The scleral coil voltage induced by those fields is decomposed into 3 signals, each related to the coil's orientation relative to the axis of one field component. Unlike older systems based on analog demodulation and filtering, this system uses a field-programmable gate array (FPGA) to oversample each induced scleral coil voltage (at 25 Msamples/s), demodulate in the digital domain, and average over 25 ksamples per data point to generate 1 ksamples/s output in real time. RESULTS: Noise floor is <0.036° peak-to-peak and linearity error is < 0.1° during 345° rotations in all three dimensions. CONCLUSION AND SIGNIFICANCE: This FPGA-based design, which is both reprogrammable and freely available upon request, delivers sufficient performance to record eye movements at high spatial and temporal precision and accuracy using coils small enough for use with small animals.