Simultaneous Intra- and Extracochlear Electrocochleography During Electrode Insertion
33
Citations
23
References
2020
Year
Signals in intracochlear ECochG recordings are reliably larger than ECochG signals recorded simultaneously from an extracochlear location. Intracochlear ECochG recordings show reversible amplitude and phase changes during insertion, not reflected in simultaneous extracochlear ECochG recordings. Such changes are most likely due to the movement of the recording electrode in relation to the signal generators. Residual high-frequency hearing is associated with larger ECochG signal amplitudes. Modeling of expected intracochlear ECochG changes during electrode insertions may allow detection of cochlear trauma in the future.