Tumor-site specific geometric distortions in high field integrated magnetic resonance linear accelerator radiotherapy

Signe Winther Hasler, Uffe Bernchou, A. Bertelsen, Elisabeth van Veldhuizen, Tine Schytte, Vibeke N. Hansen, Carsten Brink, Faisal Mahmood

Physics and Imaging in Radiation Oncology · 2020 · 13 citations · 18 references

DOIFull text

Open access

Concepts

Abstract

Magnetic resonance imaging (MRI) has exquisite soft-tissue contrast and is the foundation for image guided radiotherapy (IGRT) with integrated magnetic resonance linacs. However, MRI suffers from geometrical distortions. In this study the MRI system- and patient-induced geometric distortion at four different tumor-sites was investigated: adrenal gland (7 patients), liver (4 patients), pancreas (6 patients), prostate (20 patients). Maximum level of total distortion within the gross-tumor-volume (GTV) was 0.96 mm with no significant difference between abdominal patients (adrenal gland, liver, pancreas) and pelvic patients (prostate). Total tumor-site specific distortion depended on location in the field-of-view and increased with the distance to MRI iso-center.

References

18