OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · 1989 · 49 citations · 0 references
PharmacyChemistryAnalytical ChemistryChromatographic AnalysisMass Spectrometric AnalysisAnalytical ChromatographyChromatographyExtraction TechniquesGas ChromatographyInstrumental AnalysisAnalytical Methods DevelopmentAnalytical Test DataExtraction Of Organic CompoundsGas Chromatography-mass SpectrometryInterlaboratory ComparisonsInterlaboratory Method ComparisonInterlaboratory StudyMethod Study
An interlaboratory collaborative study was conducted to determine the precision and bias (recovery) of Office of Solid Waste (OSW) 8270/3510 for the analysis of 59 semivolatile organic compounds in groundwater. OSW Method 8270 is entitled 'GC/MS Method for Semivolatile Organics Capillary Column Technique' and includes instructions for quality control, sample preparation, and analysis of samples by GC/MS. OSW Method 3510 is entitled 'Separatory Funnel Liqui-Liquid Extraction.' The study design was based upon Youden's non-replicate plan for collaborative tests of analytical methods. Each water type was spiked with three Youden sample pairs of the 59 semivolatile compounds and analyzed using Methods 8270/3510. The test waters included reagent water as a 'control' against which other test data were compared and two types of ground water. The resulting data were analyzed using the USEPA computer program entitled, 'Interlaboratory Method Validation Study (IMVS).' The data analyses produced measures of precision and recovery for each compound in each water type and compared the performance of the method between water types.