Medical Physics · 2020 · 10 citations · 17 references
MedicineOncologyRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBreast Cancer ImagingComputed TomographyMedical ImagingRadiographic ImagingX-ray OpticsBreast X-ray ImagingDark-field ImagingLobular Carcinoma In SituRadiologic ImagingSpatial Resolution MeasurementX-ray CollimatorsX-ray Imaging PhysicsX-ray Imaging Systems
PURPOSE: This paper describes and experimentally validates a methodology for improving contrast and spatial resolution of the x-ray dark-field imaging (XDFI) by cutting the monochromator-collimator asymmetrically and thinning the Laue angle analyzer. METHODS: We measure the spatial resolution of our XDFI setup using a test object consisting of wolfram tungsten meshes and compare it with the theoretical prediction. Using x-ray dark-field computed tomography of breast cancer specimens (lobular carcinoma in situ), we demonstrate that the resolution of XDFI is sufficient for histopathologic analysis. RESULTS: Our experimental results show that the overall spatial resolution of XDFI can be improved by approximately a factor of 2 when these modifications are implemented. The reconstructed images of breast cancer specimens provide sufficient details for radiologic histopathology. CONCLUSIONS: By cutting the monochromator-collimator and thinning the Laue angle analyzer, XDFI can achieve the resolution sufficient for radiologic histopathology.