Finish before the start: Analyzing preanalytical sample errors in a tertiary care hematology laboratory
29
Citations
13
References
2020
Year
A total of 189,104 samples were received in the clinical hematology laboratory during the aforementioned period, out of which preanalytical errors were found in 4052 (2.14%) samples. Inadequate sample quantity (ISQ) comprised the bulk of preanalytical errors in our laboratory (1.11% of total samples) followed by sample clots (0.88%). There was no significant difference in the error frequencies in OPD and IPD (P = 0.1031). The proportion of errors was higher in routine vis-à -vis emergency samples and also in samples sent for coagulation analysis vis-à -vis CBC.