Analytical Chemistry · 2020 · 20 citations · 12 references
Natural SciencesChemistryAnalytical ChemistryMass Spectrometric AnalysisSensorsDetector SystemsIonization ProcessesMass SpectrometryAnalytical Detection LimitsGas Detection LimitsGatingIon Mobility SpectrometryIonizationKetonesPolaritiesTrace Gas Detection
Ion mobility spectrometers (IMS) with field switching ion shutters are an excellent choice for trace gas detection, being extremely sensitive while having fast response times. However, as different target molecules may form positive, negative, or even ions of both polarities, it is beneficial to simultaneously detect both ion polarities. Here, we present a dual drift tube IMS with a new dual field switching ion shutter for gating both ion polarities and an X-ray ionization source in orthogonal configuration. The dual field switching ion shutter allows significantly improved ion gating and ion accumulation due to improved shielding of the ionization region from the drift field. Equipped with two 75 mm long high-performance drift tubes, the dual IMS reaches high resolving power of R = 90 with detection limits in the lower pptv range for different ketones, chlorinated hydrocarbons and methyl salicylate that forms ions in both polarities.
12
Absolute Values of the Electron Mobility in Hydrogen
Norris E. Bradbury, Russell A. Nielsen · Physical Review · 1936 · 339 citations
A compact high resolution ion mobility spectrometer for fast trace gas analysis
Ansgar T. Kirk, Maria Allers, Philipp Cochems et al. · The Analyst · 2013 · 109 citations · Full text
Carolin Drees, Wolfgang Vautz, Sascha Liedtke et al. · Applied Microbiology and Biotechnology · 2019 · 60 citations · Full text