Biomechanical Analysis of Plate Fixation Compared With Various Screw Configurations for Use in the Latarjet Procedure

Rachel M. Frank, Martina Roth, Coen A. Wijdicks, Nicole Fischer, Alberto Costantini, Giovanni Di Giacomo, Anthony A. Romeo

Orthopaedic Journal of Sports Medicine · 2020 · 17 citations · 21 references

DOIFull text

Open access

Concepts

Abstract

Background: The biomechanical properties of coracoid fixation with a miniplate during the Latarjet procedure have not been described. Purpose: To determine the biomechanical properties of miniplate fixation for the Latarjet procedure compared with various screw fixation configurations. Study Design: Controlled laboratory study. Methods: A total of 8 groups (n = 5 specimens per group) were tested at a screw insertion angle of 0°: (1) 3.75-mm single screw, (2) 3.75-mm double screw, (3) 3.75-mm double screw with washers, (4) 3.75-mm double screw with a miniplate, (5) 4.00-mm single screw, (6) 4.00-mm double screw, (7) 4.00-mm double screw with washers, and (8) 4.00-mm double screw with a miniplate. In addition, similar to groups 1 to 3 and 5 to 7, there were 30 additional specimens (n = 5 per group) tested at a screw insertion angle of 15° (groups 9-14). To maintain specimen uniformity, rigid polyurethane foam blocks were used. Testing parameters included a preload of 214 N for 10 seconds, cyclical loading from 184 to 736 N at 1 Hz for 100 cycles, and failure loading at a rate of 15 mm/min until 10 mm of displacement or specimen failure occurred. Results: All single-screw constructs and 77% of 15° screw constructs failed before the completion of cyclical loading. Across all groups, group 8 (4.00-mm double screw with miniplate) demonstrated the highest maximum failure load ( P < .001). There were no differences in failure loads among specimens with single-screw fixation (groups 1, 5, 9, and 12; P > .05). All specimens in groups 9, 10, 11, 12, 13, and 14 (insertion angle of 15°) had significantly lower maximum failure loads compared with specimens in groups 2, 3, 4, 6, 7, and 8 (insertion angle of 0°) ( P < .001 for all). Conclusion: These results indicate significantly superior failure loads with the miniplate compared with all other constructs. Across all fixation techniques and screw sizes, constructs with screws inserted at 0° performed better than constructs with screws inserted at 15°. Clinical Relevance: The use of a miniplate for coracoid fixation during the Latarjet procedure may provide a more durable construct for the high-demand contact athlete.

References

21