Publication | Open Access
Optimal local impedance drops for an effective radiofrequency ablation during cavo‐tricuspid isthmus ablation
22
Citations
18
References
2020
Year
During the LI-guided CTI ablation, the effective RF ablation sites exhibited significantly greater absolute and percentage LI drops than the ineffective RF ablation sites. Absolute and percentage LI drops of 12 ohms and 11.6% may be suitable targets for effective ablation.