Passive UHF RFID Voice Prosthesis Mounted Sensor for Microbial Growth Detection

Viktorija Makarovaite, Aaron J. R. Hillier, Simon J. Holder, Campbell W. Gourlay, John C. Batchelor

IEEE Journal of Radio Frequency Identification · 2020 · 10 citations · 11 references

DOIFull text

Open access

Concepts

Abstract

Capacitive loading due to human tissue can lead to low efficiency for implantable Passive Radio Frequency Identification (RFID) antennas. The presented passive UHF antenna sensor provides read distances above 0.5 meters (within a body phantom) by utilizing a convoluted half-wave dipole design. It is able to detect simulated early to mature Candida albicans biofilm growth when mounted upon a voice prosthesis (up to a 30 μm biofilm thickness). Depending on the propagation frequency of interest, as early 4-hour growth (5 to 10 μm biofilm thickness) equivalent could be detected and before any device failure could occur due to the colonization. This was accomplished by utilising thin layers of polyurethane to decouple the saliva from the presented UHF sensor (biofilm growth is known to increase layer hydrophobicity). This presented sensor has better functionality within the U.S. UHF frequency band as it detects changes above 5 μm. If there is a need for implantation within additional tissues with variable dielectric properties, a shunt capacitance of 2.6 pF could allow the system functionality within the permittivity range of 21 to 58. Allowing for immediate medical intervention before medical prosthesis failure.