IEEE Transactions on Instrumentation and Measurement · 2020 · 28 citations · 15 references
EngineeringElectrical EngineeringInstrumentationUltrasonographyUltrasoundDiagnostic UltrasonographyFlow MetrologyTransducer TechnologyUltrasonicsUltrasound InstrumentationUltrasound PhysicsSignal-to-noise RatioTransit-time Flow MeasurementUltrasonic Flow MeasurementUltrasonic Flow SensorsUltrasonic FlowmetersUltrasonic SensorsUltrasonic WavesZero-crossing Detection
Transit-time ultrasonic flowmeters are widely used in industry to measure fluid flow. In practice, ultrasonic flowmeters either show a zero-flow error or suffer from a significant random error due to a limited signal-to-noise ratio, requiring a significant amount of averaging to achieve good precision. This work presents a method that minimizes the zero-flow error while keeping the random error low, independent of the hardware used. The proposed algorithm can adjust to changing zero-flow errors, while a flow is present. The technique combines the benefits of two common methods of determining the transit time difference between the upstream and downstream ultrasonic waves: cross correlation and zero-crossing detection. The algorithm is verified experimentally using a flow loop. It is shown that the zero-flow error can be greatly reduced without compromising the random error or increasing circuit complexity.
15
Multipath ultrasonic gas flow-meter based on multiple reference waves
Hongliang Zhou, Tao Ji, Ruichen Wang et al. · Ultrasonics · 2017 · 52 citations · Full text
Natural Sciences, Physics, Acoustics +13