A simple and reliable analytical method for simultaneous quantification of first line antitubercular drugs in human plasma by LCMS/MS

Bijoy Kumar Panda, Medha Bargaje

Analytical Methods · 2020 · 10 citations · 24 references

Concepts

Abstract

for Z were obtained. The intra-day and inter-day accuracy and imprecision were within CV ±15%. The use of the conventional silica column in the extraction of these compounds through a single step protein precipitation method simplifies the analytical process. In addition due to its simplicity and sensitivity, it can be applied to carry out therapeutic drug monitoring and drug level assessment in human plasma samples. The results of analyte levels are comparable to other reported methods. The method had been successfully applied for simultaneous determination of first line anti-tubercular drugs in pulmonary tubercular patients. The method requires 150 μL of patient plasma and offers low volume for injection (10 μL) and blood sample collection (3 mL) which will be an added advantage for pediatric anti-TB drug level assessment and monitoring.

References

24