Concepedia

Publication | Open Access

The Impact of Torso Signal Processing on Noninvasive Electrocardiographic Imaging Reconstructions

25

Citations

18

References

2020

Year

Abstract

BDR improves reconstructed electrogram morphologies and amplitudes due to a reduction in lambda value selected for the inverse problem. The simplest method (resetting the isoelectric point) is sufficient to see these improvements. HFR does not impact electrogram accuracy, but does impact post-processing to extract features such as activation times. Removal of line noise is insufficient to see these changes. HFR should be applied post-reconstruction to ensure over-filtering does not occur.

References

YearCitations