Photon-counting computed tomography of lanthanide contrast agents with a high-flux 330-μm-pitch cadmium zinc telluride detector in a table-top system

Chelsea A. S. Dunning, J. William O’Connell, Spencer M. Robinson, Kevin Murphy, Adriaan L. Frencken, Frank C. J. M. van Veggel, Kris Iniewski, Magdalena Bazalova‐Carter

Journal of Medical Imaging · 2020 · 16 citations · 26 references

DOIFull text

Open access

Concepts

Abstract

Purpose: We present photon-counting computed tomography (PCCT) imaging of contrast agent triplets similar in atomic number (Z) achieved with a high-flux cadmium zinc telluride (CZT) detector. Approach: The table-top PCCT imaging system included a 330-μm-pitch CZT detector of size 8 mm × 24 mm2 capable of using six energy bins. Four 3D-printed 3-cm-diameter phantoms each contained seven 6-mm-diameter vials with water and low and high concentration solutions of various contrast agents. Lanthanum (Z = 57), gadolinium (Gd) (Z = 64), and lutetium (Z = 71) were imaged together and so were iodine (Z = 53), Gd, and holmium (Z = 67). Each phantom was imaged with 1-mm aluminum-filtered 120-kVp cone beam x rays to produce six energy-binned computed tomography (CT) images. Results:K-edge images were reconstructed using a weighted sum of six CT images, which distinguished each contrast agent with a root-mean-square error (RMSE) of <0.29 % and 0.51% for the 0.5% and 5% concentrations, respectively. Minimal cross-contamination in each K-edge image was seen, with RMSE values <0.27 % in vials with no contrast. Conclusion: This is the first preliminary demonstration of simultaneously imaging three similar Z contrast agents with a difference in Z as low as 3.

References

26