IEEE Transactions on Biomedical Circuits and Systems · 2020 · 56 citations · 28 references
EngineeringBioengineeringBiomedical EngineeringAnesthesiaBioinstrumentationNeural EngineeringAnesthetic MonitoringElectroencephalogram AnalysisElectroencephalographic MonitoringElectroencephalographyIntraoperative MonitoringNeural Signal AcquisitionAnesthesia MonitoringConvolutional Neural NetworksDeep Learning For Eeg AnalysisDeep Learning For ElectroencephalogramElectroencephalogram ModelingElectroencephalographic Depth MonitoringElectroencephalography InstrumentationInput-referred Noise
In this article, we present a real-time electroencephalogram (EEG) based depth of anesthesia (DoA) monitoring system in conjunction with a deep learning framework, AnesNET. An EEG analog front-end (AFE) that can compensate ±380-mV electrode DC offset using a coarse digital DC servo loop is implemented in the proposed system. The EEG-based MAC, EEGMAC, is introduced as a novel index to accurately predict the DoA, which is designed for applying to patients anesthetized by both volatile and intravenous agents. The proposed deep learning protocol consists of four layers of convolutional neural network and two dense layers. In addition, we optimize the complexity of the deep neural network (DNN) to operate on a microcomputer such as the Raspberry Pi 3, realizing a cost-effective small-size DoA monitoring system. Fabricated in 110-nm CMOS, the prototype AFE consumes 4.33 μW per channel and has the input-referred noise of 0.29 μVrms from 0.5 to 100 Hz with the noise efficiency factor of 2.2. The proposed DNN was evaluated with pre-recorded EEG data from 374 subjects administrated by inhalational anesthetics under surgery, achieving an average squared and absolute errors of 0.048 and 0.05, respectively. The EEGMAC with subjects anesthetized by an intravenous agent also showed a good agreement with the bispectral index value, confirming the proposed DoA index is applicable to both anesthetics. The implemented monitoring system with the Raspberry Pi 3 estimates the EEGMAC within 20 ms, which is about thousand-fold faster than the BIS estimation in literature.
28
Adam: A Method for Stochastic Optimization
Diederik P. Kingma, Jimmy Ba · UvA-DARE (University of Amsterdam) · 2014 · 84.8K citations · Full text
Dropout: a simple way to prevent neural networks from overfitting
Nitish Srivastava, Geoffrey E. Hinton, Alex Krizhevsky et al. · 2014 · 34.3K citations
A Concordance Correlation Coefficient to Evaluate Reproducibility
Biometrics · 1989 · 8.7K citations
Qinwen Fan, Fabio Sebastiano, Johan H. Huijsing et al. · IEEE Journal of Solid-State Circuits · 2011 · 446 citations
Engineering, Biomedical Engineering, Electrical Engineering +17