The Journal of Advanced Prosthodontics · 2020 · 27 citations · 25 references
DentistryImplant DentistryOrthopaedic SurgeryBone AugmentationImplant SurgeryOrthopedicsImplant DesignImplant Site AssessmentBiomechanical Testing Of ImplantsHealing Abutment AssessmentHealing Abutment EvaluationImplant ReliabilityImplant StabilityMaxillofacial ImplantsPeriotest MeasurementStability Test Procedure
PURPOSE: study was to examine the reliability of the Anycheck device and the effect of the healing abutment diameter on the Anycheck values (implant stability test, IST). MATERIALS AND METHODS: Thirty implants were placed into three artificial bone blocks with 10 Ncm, 15 Ncm, and 35 Ncm insertion torque value (ITV), respectively (n = 10). (1) The implant stability was measured with three different kinds of devices (Periotest M, Osstell ISQ Mentor, and Anycheck). (2) Five different diameters (4.0, 4.5, 4.8, 5.5, and 6.0 mm) of healing abutments of the same height were connected to the implants and the implant stability was measured four times in different directions with Anycheck. The measured mean values were statistically analyzed. RESULTS: <.01). There were no statistically significant differences among the IST values with different healing abutment diameters. CONCLUSION: There was a strong correlation between the Periotest M and Anycheck values and between the ISQ and IST. The diameter of the healing abutment had no effect on the Anycheck values.
25
Assessment of implant stability as a prognostic determinant.
Neil Meredith · PubMed · 1999 · 822 citations
Diagnostic parameters for monitoring peri-implant conditions.
Giovanni E. Salvi, Niklaus P. Lang · PubMed · 2004 · 298 citations
Resonance frequency analysis measurements of implants at placement surgery.
Pär‐Olov Östman, Mats Hellman, Inger Wendelhag et al. · PubMed · 2006 · 259 citations