Automatic Derivation of Continuous Positive Airway Pressure Settings: Comparison with In-Laboratory Titration

Hyeyun Kim, Mi-Ja Lee, Young Hwangbo, Kwang Ik Yang

Journal of Clinical Neurology · 2020 · 12 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND AND PURPOSE: This study was designed to investigate differences in the final recommended pressure setting between that derived from an autotitrating continuous positive airway pressure (APAP) device and manual in-laboratory continuous positive airway pressure (CPAP) titration, as well as the factors that influence pressure differences in patients with obstructive sleep apnea (OSA). METHODS: + (ΔP ≤-1), and 3) APAP₉₀+ (ΔP ≥1). Regression analysis was conducted to identify predictive factors associated with ΔP. RESULTS: +, and APAP₉₀+ groups comprised 9 (18%), 23 (46%), and 18 (36%) subjects, respectively. Regression analyses revealed that male sex [β=3.539, 95% confidence interval (CI)=0.040-7.039], body mass index (BMI) (β=0.186, 95% CI=0.020-0.352), and average usage per day (β=0.768, 95% CI=0.077-1.459) were associated with ΔP. CONCLUSIONS: , there was a discordance majority showing different single pressures obtained when applying the two titration methods. Being Male, having an increased BMI, and having an increased average usage per day of APAP were significantly correlated with increased ΔP in this study.