RSC Advances · 2020 · 116 citations · 42 references
EngineeringMaterials EngineeringMaterials ScienceBiosensorsNanomaterialsSensorsChemical SensorsElectrochemical BiosensorsElectrochemical DetectionElectrochemical SensorsNanomaterial-based BiosensorsNanosensorsChitosanGlassy Carbon ElectrodesMolecularly Imprinted Polymer SensorsNanomaterial-based SensorsNanomaterial-modified ElectrodesNanoparticlesSelectivitySensing
. The synergistic combination of Au nanoparticles and Ch-MIP helps detect the CIP antibiotic with good sensitivity and selectivity, respectively. We investigated the selectivity aspect by using some possible interfering species and the developed sensing system showed good selectivity for CIP with a 66% response compared to the other compounds (≤45% response). The proposed sensing strategy showed its applicability for successful detection of CIP in real samples like tap water, mineral water, milk, and pharmaceutical formulation. The developed sensor showed good selectivity towards CIP even among the analogue molecules of Norfloxacin (NFX) and Ofloxacin (OFX). The developed sensor was successfully applied to determine the CIP in different samples with a satisfactory recovery in the range of 94 to 106%.
42
Xing Xuan, Hyo Sang Yoon, Jae Yeong Park · Biosensors and Bioelectronics · 2018 · 410 citations
Dhana Lakshmi, Alessandra Bossi, Michael J. Whitcombe et al. · Analytical Chemistry · 2009 · 380 citations