Changing the syringe pump: A challenging procedure in critically ill patients
17
Citations
13
References
2020
Year
The syringe change can be a very critical moment given different influencing variables. Syringe pump position, displaced higher than the patient level, always generates a medication bolus that is higher at the lowering of the central venous pressure value. The presence of a neutral displacement needle-free connector reduces the incidence of boluses. When the pump is placed at the patient level, the presence of neutral displacement needle-free connector reduces the establishment of boluses, even in a central venous pressure of -5 mmHg simulations.