Technical Note: Validation of TG 233 phantom methodology to characterize noise and dose in patient CT data
25
Citations
16
References
2020
Year
The Mercury Phantom provided a close prediction of radiation dose and image noise in clinical patient images. By assessing dose and image quality in a phantom with multiple sizes, protocol parameters can be designed and optimized per patient size in a highly constrained setup to predict clinical scanner and ATCM system performance.