Publication | Closed Access
Prediction of radiation‐induced malfunction for cardiac implantable electronic devices (CIEDs)
21
Citations
59
References
2020
Year
Numerical understanding of the malfunction of CIEDs has been attained for the first time. It has been clarified that neutron dose is a good scale for the risk of CIEDs in radiotherapy. Prediction of the frequency of malfunction as well as discussion of the risk to CIEDs in radiotherapy among the multiple modalities have become possible. Because the present study quantitatively clarifies the neutron contribution to malfunction, revision of clinical guidelines is suggested.