IEEE Solid-State Circuits Letters · 2019 · 56 citations · 6 references
EngineeringBiomedical EngineeringElectrical EngineeringAnalog ElectronicsOptoelectronicsSensorsAnalog IntegrationDynamic Power ReductionLight-emitting DiodesPhotodetectorsSensor Signal ConditioningDirect MeasurementDuty-cycled Sensor OperationDynamic RangeIntegrated PhotodetectorsOptical Readout SystemsReadoutReadout ElectronicsReadout Integrated Circuits
This letter reports on a photoplethysmogram (PPG) readout IC (ROIC), comprising an ultralow-power recording front-end and a 2-channel light-emitting diode (LED) driver. The front-end uniquely employs a continuous-time, first-order, incremental delta-sigma modulator (I-AΣM) for direct light-to-digital conversion in a single stage with duty-cycled operation for power savings. Moreover, the I-AΣM enables reconfigurable operation in intermittent conversion phases that allows the direct measurement of dc and ambient light-induced components of the photodiode current for signal-aware on-the-fly cancellation to recover the small ac PPG signal. Fabricated in 0.18-μm 1P/6M CMOS, the PPG ROIC features a high dynamic range of 92.7 dB, dissipates 8.1 μW and 1.95 mW in the recording front-end and the LED (with its driver), respectively, with ~10% duty cycling, and is validated by high-fidelity PPG recording from a human subject fingertip using an infrared LED.
6